SHIMADZU SIMULTANEOUS TGA/DTA ANALYZER DTG-60AH PRODUCT SPECIFICATION
The Shimadzu automatic simultaneous thermal analyzer DTG-60AH provides simultaneous thermogravimetry/derivative thermogravimetry (TGA/DTG) and differential thermal analysis (DTA) on a single sample.
A high resolution top loading thermobalance provides TGA data along with two detectors for semi-quantitative DTA/DSC data. A motorized, self locating bifilar wound furnace is used. A 24 position autosampler including punching mechanism is built-in.
Temperature range is ambient to 1500°C.
Included in the base instrument:
- Two plug-in detectors and set of crucibles (aluminium, platinum)
- Lighted LCD display for temperature and signal
- Operation panel for direct access for emergency stop, autosampler adjustment and service tasks
- Remote system functions
- Self-diagnostics by software
- Three gas connectors
- Operation manuals
- Built in cooling fan allows controlled cyclic runs between 50°C and 1500°C
- Auto-cleaning function
- Built-in autosampler
- Full control of autosampler function by optical and electronic devices
- Adjustable punch mechanism for volatile samples
| Temperature Range | RT to 1500°C | | DTA sensitivity | 0.1 µV |
| Heating rate | ±0.01 to ±99.9°K/min at 0.1°K intervals | Noise | <0.2 µV |
| Temp reproducibility | ±0.1°K | Time constant | <20 seconds |
| Temp accuracy | ±0.1°K | Crucible volume | 50-80 µl (typical) |
| Time to cool | <13 min to 100°C | Atmosphere | inert or reactive |
| Sample mass | 1000 mg | Autosampler | 24 positions (2nd tray optional) |
| TG mass range | ±500 mg | Punch mechanism | built-in, for volatile samples |
| TG noise | <3 µg | |
| TG sensitivity | 1 µg |
| TG readability | 0.1 µg |
| Dynamic range | ±1000 µV |
|